Validation of low-dose lung cancer PET-CT protocol and PET image improvement using machine learning

Authors: Nai, Y.H. et al.

Journal: Physica Medica

Volume: 81

Pages: 285-294

eISSN: 1724-191X

ISSN: 1120-1797

DOI: 10.1016/j.ejmp.2020.11.027

Abstract:

Purpose: To conduct a simplified lesion-detection task of a low-dose (LD) PET-CT protocol for frequent lung screening using 30% of the effective PETCT dose and to investigate the feasibility of increasing clinical value of low-statistics scans using machine learning. Methods: We acquired 33 SD PET images, of which 13 had actual LD (ALD) PET, and simulated LD (SLD) PET images at seven different count levels from the SD PET scans. We employed image quality transfer (IQT), a machine learning algorithm that performs patch-regression to map parameters from low-quality to high-quality images. At each count level, patches extracted from 23 pairs of SD/SLD PET images were used to train three IQT models – global linear, single tree, and random forest regressions with cubic patch sizes of 3 and 5 voxels. The models were then used to estimate SD images from LD images at each count level for 10 unseen subjects. Lesion-detection task was carried out on matched lesion-present and lesion-absent images. Results: LD PET-CT protocol yielded lesion detectability with sensitivity of 0.98 and specificity of 1. Random forest algorithm with cubic patch size of 5 allowed further 11.7% reduction in the effective PETCT dose without compromising lesion detectability, but underestimated SUV by 30%. Conclusion: LD PET-CT protocol was validated for lesion detection using ALD PET scans. Substantial image quality improvement or additional dose reduction while preserving clinical values can be achieved using machine learning methods though SUV quantification may be biased and adjustment of our research protocol is required for clinical use.

Source: Scopus

Validation of low-dose lung cancer PET-CT protocol and PET image improvement using machine learning.

Authors: Nai, Y.-H. et al.

Journal: Phys Med

Volume: 81

Pages: 285-294

eISSN: 1724-191X

DOI: 10.1016/j.ejmp.2020.11.027

Abstract:

PURPOSE: To conduct a simplified lesion-detection task of a low-dose (LD) PET-CT protocol for frequent lung screening using 30% of the effective PETCT dose and to investigate the feasibility of increasing clinical value of low-statistics scans using machine learning. METHODS: We acquired 33 SD PET images, of which 13 had actual LD (ALD) PET, and simulated LD (SLD) PET images at seven different count levels from the SD PET scans. We employed image quality transfer (IQT), a machine learning algorithm that performs patch-regression to map parameters from low-quality to high-quality images. At each count level, patches extracted from 23 pairs of SD/SLD PET images were used to train three IQT models - global linear, single tree, and random forest regressions with cubic patch sizes of 3 and 5 voxels. The models were then used to estimate SD images from LD images at each count level for 10 unseen subjects. Lesion-detection task was carried out on matched lesion-present and lesion-absent images. RESULTS: LD PET-CT protocol yielded lesion detectability with sensitivity of 0.98 and specificity of 1. Random forest algorithm with cubic patch size of 5 allowed further 11.7% reduction in the effective PETCT dose without compromising lesion detectability, but underestimated SUV by 30%. CONCLUSION: LD PET-CT protocol was validated for lesion detection using ALD PET scans. Substantial image quality improvement or additional dose reduction while preserving clinical values can be achieved using machine learning methods though SUV quantification may be biased and adjustment of our research protocol is required for clinical use.

Source: PubMed

Validation of low-dose lung cancer PET-CT protocol and PET image improvement using machine learning.

Authors: Nai, Y.-H. et al.

Journal: Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)

Volume: 81

Pages: 285-294

eISSN: 1724-191X

ISSN: 1120-1797

DOI: 10.1016/j.ejmp.2020.11.027

Abstract:

Purpose

To conduct a simplified lesion-detection task of a low-dose (LD) PET-CT protocol for frequent lung screening using 30% of the effective PETCT dose and to investigate the feasibility of increasing clinical value of low-statistics scans using machine learning.

Methods

We acquired 33 SD PET images, of which 13 had actual LD (ALD) PET, and simulated LD (SLD) PET images at seven different count levels from the SD PET scans. We employed image quality transfer (IQT), a machine learning algorithm that performs patch-regression to map parameters from low-quality to high-quality images. At each count level, patches extracted from 23 pairs of SD/SLD PET images were used to train three IQT models - global linear, single tree, and random forest regressions with cubic patch sizes of 3 and 5 voxels. The models were then used to estimate SD images from LD images at each count level for 10 unseen subjects. Lesion-detection task was carried out on matched lesion-present and lesion-absent images.

Results

LD PET-CT protocol yielded lesion detectability with sensitivity of 0.98 and specificity of 1. Random forest algorithm with cubic patch size of 5 allowed further 11.7% reduction in the effective PETCT dose without compromising lesion detectability, but underestimated SUV by 30%.

Conclusion

LD PET-CT protocol was validated for lesion detection using ALD PET scans. Substantial image quality improvement or additional dose reduction while preserving clinical values can be achieved using machine learning methods though SUV quantification may be biased and adjustment of our research protocol is required for clinical use.

Source: Europe PubMed Central