Abstract:

PURPOSE: Iterative reconstruction algorithms are widely used to reconstruct positron emission tomography computerised tomography (PET/CT) data. Lesion detection in the liver by 18F-fluorodeoxyglucose PET/CT (18F-FDG-PET/CT) is hindered by 18F-FDG uptake in background liver parenchyma. The aim of this study was to compare semi-quantitative parameters of histologically-proven colorectal liver metastases detected by 18F-FDG-PET/CT using data based on a Bayesian penalised likelihood (BPL) reconstruction, with data based on a conventional time-of-flight (ToF) ordered subsets expectation maximisation (OSEM) reconstruction. METHODS: A BPL reconstruction algorithm was used to retrospectively reconstruct sinogram PET data. This data was compared with OSEM reconstructions. A volume of interest was placed within normal background liver parenchyma. Lesions were segmented using automated thresholding. Lesion maximum standardised uptake value (SUVmax), standard deviation of background liver parenchyma SUV, signal-to-background ratio (SBR), and signal-to-noise ratio (SNR) were collated. Data was analysed using paired Student's t-tests and the Pearson correlation. RESULTS: Forty-two liver metastases from twenty-four patients were included in the analysis. The average lesion SUVmax increased from 8.8 to 11.6 (p<0.001) after application of the BPL algorithm, with no significant difference in background noise. SBR increased from 4.0 to 4.9 (p<0.001) and SNR increased from 10.6 to 13.1 (p<0.001) using BPL. There was a statistically significant negative correlation between lesion size and the percentage increase in lesion SUVmax (p=0.03). CONCLUSIONS: This BPL reconstruction algorithm improved SNR and SBR for colorectal liver metastases detected by 18F-FDG-PET/CT, increasing the lesion SUVmax without increasing background liver SUV or image noise. This may improve the detection of FDG-avid focal liver lesions and the diagnostic performance of clinical 18F-FDG-PET/CT in this setting, with the largest impact for small foci.

Source: PubMed

Does a novel penalized likelihood reconstruction of 18F-FDG PET-CT improve signal-to-background in colorectal liver metastases?

Authors: Parvizi, N., Franklin, J., McGowan, D., Teoh, E.J., Bradley, K.M. and Gleeson, F.V.

Journal: European Journal of Radiology

Volume: 84

Issue: 10

Pages: 1873-1878

Publisher: Elsevier

ISSN: 0720-048X

DOI: 10.1016/j.ejrad.2015.06.025

Source: Manual

Does a novel penalized likelihood reconstruction of 18F-FDG PET-CT improve signal-to-background in colorectal liver metastases?

Authors: Parvizi, N., Franklin, J.M., McGowan, D.R., Teoh, E.J., Bradley, K.M. and Gleeson, F.V.

Journal: European journal of radiology

Volume: 84

Issue: 10

Pages: 1873-1878

eISSN: 1872-7727

ISSN: 0720-048X

DOI: 10.1016/j.ejrad.2015.06.025

Abstract:

Purpose

Iterative reconstruction algorithms are widely used to reconstruct positron emission tomography computerised tomography (PET/CT) data. Lesion detection in the liver by 18F-fluorodeoxyglucose PET/CT (18F-FDG-PET/CT) is hindered by 18F-FDG uptake in background liver parenchyma. The aim of this study was to compare semi-quantitative parameters of histologically-proven colorectal liver metastases detected by 18F-FDG-PET/CT using data based on a Bayesian penalised likelihood (BPL) reconstruction, with data based on a conventional time-of-flight (ToF) ordered subsets expectation maximisation (OSEM) reconstruction.

Methods

A BPL reconstruction algorithm was used to retrospectively reconstruct sinogram PET data. This data was compared with OSEM reconstructions. A volume of interest was placed within normal background liver parenchyma. Lesions were segmented using automated thresholding. Lesion maximum standardised uptake value (SUVmax), standard deviation of background liver parenchyma SUV, signal-to-background ratio (SBR), and signal-to-noise ratio (SNR) were collated. Data was analysed using paired Student's t-tests and the Pearson correlation.

Results

Forty-two liver metastases from twenty-four patients were included in the analysis. The average lesion SUVmax increased from 8.8 to 11.6 (p<0.001) after application of the BPL algorithm, with no significant difference in background noise. SBR increased from 4.0 to 4.9 (p<0.001) and SNR increased from 10.6 to 13.1 (p<0.001) using BPL. There was a statistically significant negative correlation between lesion size and the percentage increase in lesion SUVmax (p=0.03).

Conclusions

This BPL reconstruction algorithm improved SNR and SBR for colorectal liver metastases detected by 18F-FDG-PET/CT, increasing the lesion SUVmax without increasing background liver SUV or image noise. This may improve the detection of FDG-avid focal liver lesions and the diagnostic performance of clinical 18F-FDG-PET/CT in this setting, with the largest impact for small foci.